XXRapidFrame or XX RapidFrame or XX Rapid Frame ICCD camera


ICCD framing camera system and ICCD multi-channel framing camera
gated camera and gateable camera for ultra high speed imaging ultra high speed cameras and ICCD low light cameras
 

... the eyes of your computer ...
Home Company  I  Technology Products  I  Applications  I  Downloads  I  FAQ  I  Contact 


Our Nanosecond ICCD Camera Family:  4 Quik E

Highlights



4 quik e iccd camera

Parameter
Unit
4 Quik E, digital HR
4 Quik E, digital SR
4 Quik E analog
Internal exposure time
ns
1.2ns - 80s, min steps 0.1ns (18mm image intensifier)
External exposure time
ns
1.2ns ... DC (18mm image intensifier), 1.8 ... DC (25mm image intensifier)
Trigger propagation delay

<50ns, less than 10ps jitter
Multiple exposure

any seqence, 3µs "dead time" between exposures
Dynamic range A/D
bit
14bit, up to 21 bit with 4 Spec E image and data acquisition software
with all lines integrated (binned), dynamic expansion active
Sensitivity of system

more than 1count/photoelectron pixel, up to 80s integration time on CCD
Camera output
standard: CameraLink, optional: USB 2.0
CCD output bit
12
14
12
14
EIA (Japan, USA)
CCIR (elsewhere)
PS (Progressive Scan VGA)
CCD resolution
pixel
1360 x 1024
High Resolution
(HR)
782 x 582
Standard Resolution
(SR)
768 x 494 (EIA)
752 x 582 (CCIR)

640 x 480 (PS)
CCD pixel size
µm
4.7 x 4.7
8.6 x 8.6
8.4 x 9.8 (EIA)
8.6 x 8.3 (CCIR)
9.8 x 9.8 (PS)
Frame rate (digital
fps
10.6 / 17.9 / 20.9 33.8 / 60.8 / 67.0 -
Image Frequency (analog)
Hz
-

30/60 (EIA)
25/50 (CCIR)
30 or 60 (PS)
60/120 (PS, dual scan)
Gain
db
1x1, ROI: 0..20db, 2x2: 0..25db
remote control RS232
Binning

1x1 (full frame)
2x2 (binning)
ROI (region of interest)
-
Scan mode

field/frame, selectable through computer RS232 interface
Optical interface

C-mount (standard), F-mount (optional)
Coupling lens

customized relay coupling lens between image intensifier -> CCD
no distortion, best image quality up to 180 lines/mm
Image intensifier
mm
standard: 18, image area: 14.4 x 10.8;
optional: 25, image area: 20 x 15
Photocathode

standard: S20 (UV), S25 (IR) others on request
Phosphor screen

P43 or P46
Spectral sensitivity
nm
UV  - NIR, 165 - ... - 920 (standard), 110 - ... - 1300 (optional)
Camera dimensions
mm
248 x 110 x 135 (l x w x h), without objective lens
Camera weight
kg
3, all in one head, without objective lens
Camera mount

1/4" x 20 and M8 mounting hole at the bottom of the camera
Voltage
V
12V +5% / -2%






Products
Picosecond ICCD
Nanosecond ICCD
Framing Camera
Intensifier Module
Software



Highlights








Copyright © 2008 Stanford Computer Optics, Inc.  Sitemap